Final5 CLIA Rule. Part V: Method Validation Process and Procedures

April 24, 2003

James O. Westgard, PhD

Laboratories will have to gear up to perform method validation studies for all "non-waived" methods implemented after April 24, 2003. That's today!

The Final CLIA Rule requires that moderate complexity methods be treated the same as high complexity methods, now grouped together as non-waived methods. That means method validation studies must now be performed for many methods that were previously exempted.

A review of the MV process

Here's a brief summary of our lessons on method validation. You can access each lesson via the link provided to review the material in greater detail.

Applications with published data

A critical review of the literature is always a good starting point when selecting and evaluating a method. This literature includes scientific papers as well as manufacturer's method descriptions. The time and effort needed for method validation studies in your own laboratory can be minimized by a careful assessment of the data in the literature.

Published evaluation studies seldom follow a standard validation process. Therefore, it is necessary to impose your own system of organization, data analysis, and data interpretation if you are to make sense of the published results. This is a process of critical review, which is distinctly different from just accepting the organization, data analysis, and conclusions that have been published.

Applications in your own laboratory

It is important to have a clear understanding of the method validation process and be well-organized in carrying out your experimental studies. Good record keeping is essential to document the conditions of the studies (reagent lot numbers, calibration lot numbers, re-calibrations, preventive maintenance procedures, any method changes or corrective actions).

Adaptations for individual laboratory applications

It should be recognized that each laboratory situation may be different, therefore, different adaptations are possible in different laboratories. The approach is to maintain the principles of the method validation process, while making the experimental work as efficient and practical as possible. Some ideas are presented here concerning the scope of studies, personnel skills, and data analysis techniques.

References

  1. Westgard JO. A method evaluation decision chart (MEDx) Chart for judging method performance. Clin Lab Science 1995;8:277-83. See PDF files on this website.
  2. Information for authors. Clinical Chemistry 1999;45:1-4.

See the Entire CLIA Final Rule Series:


Copyright © 2003. All rights reserved.
Westgard QC, 7614 Gray Fox Trail, Madison WI 53717
Call 608-833-4718 or e-mail us at westgard@westgard.com
A Message from JOW
QC Lessons | QC Applications | Questions | Multirule
CLIA Requirements | What's New? | Catalog | Demo Download
Home  | Glossary | ARCHIVES | Links | Feedback